MEDECO 9000D 15" Patient Monitor

Product group
Patient Monitor
Origin
China

Shanghai MedEco Industry Co., Ltd.

China
Shanghai MedEco Industry Co., Ltd.

Product description

MEDECO 9000D 15" monitor 15" color TFT screen Multi-channel display Big font interface ST and arrhythmia analysis Drug dose calculation and titration table OxyCRG dynamic view display Respiration monitoring with apnea alarms Alarm review, arrhythmia review Patient info. Input Anti-defibrillation design NIBP over pressure protection Plug-in rechargeable battery, AC/DC available. Supporting wire or wireless network.

15" color TFT screen Multi-channel display Big font interface

ST and arrhythmia analysis

Drug dose calculation and titration table

OxyCRG dynamic view display

Respiration monitoring with apnea alarms


Alarm review, arrhythmia review

Patient info. Input

Anti-defibrillation design

NIBP over pressure protection

Plug-in rechargeable battery, AC/DC available. Supporting wire or wireless network.


ECG

Lead mode 5 Leads, I, II, III, AVR, AVL, AVF, V

Gain ×2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV,auto

Heart rate Measure range:

Adult 15 ~ 300 bpm

Neonatal/Pediatric 15 ~ 350 bpm


Accuracy ±1%

Resolution 1 bpm

Sensitivity > 200 VμP-P

Differential Input Impedance > 5 M ohm


CMRR Monitor Mode >100 dB

Surgery Mode >100 dB

Diagnostic Mode >90 dB

Electrode offset potential ±300mV

Leakage Current < 10 μA

PACE pulse detect range±2~±700mV

width 0.1~2ms

rise time 10~100 μs

PACE pulse rejection range ±2 ~±700mV

width 0.1~2ms

rise time 10 ~100 μs

Baseline Recovery < 3 s After defibrillation.

Signal Range ±8 mV p-p

Bandwidth Surgery 1 ~ 20 Hz

Monitor 0.5 ~ 40 Hz

Diagnostic 0.05 ~ 130 Hz

Calibration Signal 1 mV p-p, 5% accuracy

ST measurement Range -2.0 ~+2.0 mV

Accuracy-0.8mV~+0.8mV

±0.02mV or±10%, which is greater

Other range: unspecified


Respiration

Method Impedance between RA-LL

Differential Input Impedance >2.5 MΩ

Respiration Impedance Range 0.3~3Ω

Base Impedance Range 200Ω~4000Ω

Bandwidth 0.1 ~ 2.5 Hz

Gain ×0.25,×0.500,×1,×2,×4

Respiration Rate Measurement Range

Adult 0 ~ 120 BrPM

Neonatal / Pediatric 0 ~ 150 BrPM

Resolution 1 BrPM


Accuracy 0~6 BrPM: unspecified

7~150 BrPM:±2 BrPM or±2%

use the greater

Inspiriting current < 300 A RMS max

Apnea Alarm 10 ~ 40 s


NIBP

Method Oscillometry

Measure mode Manual, Auto, Continual

Measure Interval in AUTO Mode 1,2,3,4,5,10,15,30,60,90,120,180,240,480 min

Measure Period in Continual Mode 5 min

Pulse Rate Range 40 ~ 240 bpm

Measure and Alarm Range Adult Mode

SYS 40 ~ 270 mmHg

DIA 10 ~ 215 mmHg

MEAN20 ~ 235 mmHg

Pediatric Mode

SYS 40 ~ 220 mmHg

DIA 10 ~ 150 mmHg

MEAN20 ~ 165 mmHg

Neonatal Mode

SYS 40 ~ 135 mmHg

DIA 10 ~ 100 mmHg

MEAN 20 ~ 110 mmHg

Static pressure accuracy ±3mmHg

Resolution 1mmHg

Accuracy Maximum Mean error ±5mmHg

Maximum Standard deviation 8mmHg

Overpressure Protection Adult 297±3 mmHg

Pediatric 240±3 mmHg

Neonatal 147±3 mmHg


SpO2

Measurement Range 0 ~ 100 %

Resolution 1 %

Accuracy 70% ~ 100% ±2 %

69%

0% ~ 39% unspecified


Pulse Rate Measure and Alarm Range 20~300bpm

Resolution 1bpm

Accuracy ±3bpm


TEMP

Channel 2

Measure and Alarm Range 0 ~ 50℃

Resolution 0.1℃

Accuracy(no sensor) ±0.1(25℃-45℃),0.2±℃(other)

Accuracy(include sensor) ±0.2℃(32℃-42℃)

±0.3(other)



tandard: ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature

Optional: EtCO2, IBP, Thermal Recorder