MEDECO 12.1" Color TFT Screen Monitor 9000

Product group
Patient Monitor
Origin
China

Shanghai MedEco Industry Co., Ltd.

China
Shanghai MedEco Industry Co., Ltd.

Product description

MEDECO 12.1" color TFT screen Monitor 9000 Multi‐channel display Big font interface ST and arrhythmia analysis Drug dose calculation and titration table OxyCRG dynamic view display Respiration monitoring with apnea alarms Alarm review, arrhythmia review Patient info. Input Anti‐defibrillation design NIBP over pressure protection Plug‐in rechargeable battery, AC/DC available. Supporting wire or wireless network.

Multi‐channel display

Big font interface

ST and arrhythmia analysis

Drug dose calculation and titration table

OxyCRG dynamic view display

Respiration monitoring with apnea alarms


Alarm review, arrhythmia review

Patient info. Input

Anti‐defibrillation design

NIBP over pressure protection

Plug‐in rechargeable battery, AC/DC available.

Supporting wire or wireless network.


Technical Specifications


ECG


Lead Mode 5 Leads (R, L, F, N, C or RA, LA, LL, RL,V)

Lead selection I, II, III, avR, avL, avF, V

Wavef orm 2 ch

Lead mode 3 Leads ( R, L, F or RA, LA, LL)

Lead selection I, II, III,

Wave form 1 ch

Gain 2.5mm/mV, 5.0mm/mV, 10mm/mV,20mm/mV, auto


HR and Alarm Range Adult 15 ~ 300 bpm

Neo/Ped 15 ~ 350 bpm

Accuracy ±1% or±1bpm, which is greater

Resolution 1 bpm

Sensitivity > 200 (uVp-p)

Differential Input Impedance > 5 MΩ


CMRR Monitor > 105 dB

Operation > 105 dB

Diagnosis > 85 dB

Electrode offset potential ±300mV

Leakage Current < 10 uA

Baseline Recovery < 3 S After Defi.

ECG Signal Range ±8 m V ( Vp-p )


Bandwidth Surgery 1 ~ 20 Hz

Monitor 0.5 ~ 40 Hz

Diagnostic 0.05 ~ 130 Hz

Calibration Signal 1 (mVp-p), Accuracy:5%

ST Segment Monitoring Range Measure and Alarm -2.0 ~ +2.0 mV


ARR Detecting Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,BRADY, MISSED BEATS,

PNP, PNC,Alarm Available,Review Available


Respiration


Method Impedance between R-F(RA-LL)

Differential Input Impedance > 2.5 MΩ

Measuring Impedance Range 0.3~3 Ω

Base line Impedance Range 200~4 KΩ

Bandwidth 0.1~2.5Hz

Resp. Rate Measuring and Alarm Range

Adult 0 ~ 120 Brpm

Neo/Ped 0 ~ 150 Brpm

Resolution 1 Brpm

Accuracy ±2 Brpm


Apnea Alarm 10 ~ 40 S


NIBP


Method Oscillometric

Mode Manual, Auto, STAT

Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120, 180, 240,480 (Min)

Measuring Period in STAT Mode 5 Min

Pulse Rate Range 40 ~ 240 bpm

Alarm Type SYS, DIA, MEAN


Measuring and alarm range Adult Mode

SYS 40 ~ 270 mmHg

DIA 10 ~ 215 mmHg

MEAN 20 ~ 235 mmHg

Pediatric Mode

SYS 40 ~ 200 mmHg

DIA 10 ~ 150 mmHg

MEAN 20 ~ 165 mmHg

Neonatal Mode

SYS 40 ~ 135 mmHg

DIA 10 ~ 100 mmHg

MEAN 20 ~ 110 mmHg

Resolution 1mmHg


Accuracy Maximum Mean error ±5mmHg

Maximum Standard deviation ±8mmHg

Overpressure Protection Adult Mode 297±3 mmHg

Pediatric Mode 240±3 mmHg

Neonatal Mode 147±3 mmHg


SpO2

Measuring Range 0 ~ 100 %

Alarm Range 0 ~ 100 %

Resolution 1 %

Accuracy 70% ~ 100% 2 %

0% ~ 69% unspecified

Actualization intervalabout 1 Sec.

Alarm Delay 10 Sec.

Pulse Rate Measuring and Alarm Range 20~300 bpm

Resolution 1bpm

Accuracy ±3bpm



TEMP


Channel 2

Measuring and Alarm Range 0 ~ 50 ℃

Resolution 0.1 ℃

Accuracy ±0.2 ℃

Actualization intervalabout 1 Sec.

Average Time Constant < 10 Sec.


IBP


Label ART, PA, CVP, RAP, LAP, ICP, P1, P2

Measuring and alarm range ART 0 ~ 300 mmHg

PA -6 ~ 120 mmHg

CVP/RAP/LAP/ICP-10 ~ 40 mmHg

P1 / P2-10 ~ 300 mmHg


Press Sensor Sensitivity 5 uV/V/mmHg

Impedance 300-3000Ω


Resolution 1 mmHg

Accuracy ±2% or 1mmHg which is greater

Actualization interval about 1 secretary

Supplier info

Business type
Manufacturer
Location
China
Exports to
Worldwide