Fully Automatic Biochemistry Analyzer BA-A-120

Category Fully Auto Biochemistry Analyzer
Model BA-A-120
Country of origin China

Seller information

Visit store
Bioevopeak Co Ltd
Rm. 2014, Bldg. 3, Ligaoguojihuayuan, No. 1222, West Aoti Road, Lixia District, Jinan, Shandong, China
+86-531-88982330
Manufacturer
Established 2010

Description

Biochemistry analyzer is used in a lab to check cholesterol, sugar, enzyme, protein, etc. by biological samples such as blood or urine. BA-A-120 Fully Auto Biochemistry Analyzer easy to operation with a throughput of 120 tests per hour

• Constant 120 test / hour with double reagent
• 80 reagent positions
• Automatic liquid level sensing, anti-collision protection.
• Integrated full closed optical system, multi wavelength simultaneous detection.
• New amorphous optical plastics reaction cups, perfect transmittance.
• Maintenance free semiconductor solid state refrigeration
• Friendly surface and easy to operation

Multifunctional precision sampling probe
• External & internal mirror polish, external & internal probe washing
• Dedicated sampling probe equipped with sensitive liquid sensor, timely feedback of reagent & sample residuals
• Collision protection, automatic probe depth adjusting

Excellent mixer design
• Teflon coating mixer. No water dropping(reducing carry-over)
• Excellent mixing effect with standard mixing procedure


High precision ceramic syringe
• Permanent ceramic piston
• Accurate dispensing as low as 0.1µL

Reagent tray & Sample tray;
• 80 reagent positions,40 sample positions
• Non-stop cooling system with peltier pad inside, 24 hours 2℃-14℃
• Bar code reader(optional)

Reaction tray
• 50 reaction cuvettes
• Reaction volume as low as 150μL
• Stable & accurate temperature (37±0.1℃)for reaction

Advanced operating system
• User friendly software interface simple, easy-to-operation
• Sample automatic dilution(decrease, increase, normal)
• Reaction result automatic checking and warning, calibration result automatic checking, warning alarm log
• Linear & nonlinear calibration, Multi-points calibration up to 8 points
• Quality control apply to westgard, L-T plot, cumulative

SPECIFICATIONS

General information:
Machine type: Random access, open reagent & locked reagent (optional)
Test speed: Constant 120 tests/hour (Mono reagent or double reagent)
Test principle: Colorimetric method, turbidimetry
Test method: 1 point end, 2 point end, Fixed-Time, Kinetics
Calibration type: Linear & nonlinear

Sample unit:
Sample tray: 40 sample positions
Sample tube: Micro cup & test tube & blood collection tube can be used. (Φ 12 -13 )mm *(25-100 )mm
Sample volume: 2-30uL step by 0.1uL
Sample probe: Liquid surface detection, timely tracking for liquid volume. Collision avoidance
Sample barcode: Sample barcode reader (optional)

Reagent unit:
Reagent tray: 80 reagent positions (40 for Rl & 40 for R2), Non-stop cooling system with peltier pad inside, 24 hours 2°C-14°C, open system, could accept any reagent
Reagent volume: 20-300uL, step by luL
Reagent needle: Liquid surface detection, timely tracking for liquid volume. Collision avoidance
Regent barcode: Regent barcode reader (optional)

Reaction unit:
Reaction tray: 50 reaction cuvettes, new amorphous optical plastics reaction cups, perfect transmittance.
Reaction volume: 150uL-330uL
Reaction time: 10 minutes
Temperature: Peltier pad incubation system, reaction temperature (37°C ±0.1 °C)

Mixing system:
Mixer: paddle-type mixing, effective reducing of carry-over

Optical system:
Light source: Halogen lamp
Wavelength: 340nm; 405nm; 450nm; 505nm; 546nm; 578nm; 630nm;700nm(4 more options)
Absorbance range: 0-4.0 Abs
Resolution: 0.0001 Abs

Operating system:
Operation system: Windows XP, Windows 7, Windows 8, Windows 10
Data processing: Automatic calibration, test combination, effective time management of reagents, serum index, whole process detection, linear expansion of enzyme, blank deduction,dirty cuvette memory, cross infection prevention, patient information memory and association input, automatic report audit, data fuzzy query, report statistics and printing, reference range grading, alarm information classification, user operation privilege classification management.
Report printing: 6 formats optional, support custom pattern
interface: TCP/IP network interface

Working conditions:
Power supply: AC 110/220V ± 10%, 50/60 Hz, 650W
Water consumption: 6L/hour(maximum)
Dimension: 750mm*550mm *490mm(L*W*H)
Weight: 70kg
language: All languages

Reviews

0.0 / 5
0 Reviews
5
(0)
4
(0)
3
(0)
2
(0)
1
(0)
Review this item